Basic Module: Fortron Source FSP400M-80PA
| General information | ||||
| SKU / Item Number | FSP400M-80PA | |||
| Product Basic Data | ||||
| Manufacturer | Fortron Source | |||
| Power Supply | ||||
| Typ (Input) | AC | Range | 90~264V | |
| Input frequency | 47 to 63 Hz | |||
| Connector | C14 | |||
| Power supply - Input | AC: 90~264V | |||
| Output wattage | 400 W | |||
| Output Voltage | +5V | Output Current Min. | 0,1 A | |
| Output Current Max | 18 A | |||
| Regulation Line (±) | 5 % | |||
| Output Ripple & Noise Max | 50 mV | |||
| -12V | Output Current Min. | 0 A | ||
| Output Current Max | 0,5 A | |||
| Regulation Line (±) | 5 % | |||
| Output Ripple & Noise Max | 120 mV | |||
| +3.3V | Output Current Min. | 0,1 A | ||
| Output Current Max | 18 A | |||
| Regulation Line (±) | 5 % | |||
| Output Ripple & Noise Max | 50 mV | |||
| +5V Standby | Output Current Min. | 0 A | ||
| Output Current Max | 3 A | |||
| Regulation Line (±) | 5 % | |||
| Output Ripple & Noise Max | 50 mV | |||
| -5V | Output Current Min. | 0 A | ||
| Output Current Max | 0,2 A | |||
| Regulation Line (±) | 5 % | |||
| Output Ripple & Noise Max | 100 mV | |||
| +12V1 | Output Current Min. | 0,2 A | ||
| Output Current Max | 16 A | |||
| Regulation Line (±) | 5 % | |||
| Output Ripple & Noise Max | 120 mV | |||
| +12V2 | Output Current Min. | 0,05 A | ||
| Output Current Max | 16 A | |||
| Regulation Line (±) | 5 % | |||
| Output Ripple & Noise Max | 120 mV | |||
| +12V3 | Output Current Min. | 0,1 A | ||
| Output Current Max | 16 A | |||
| Regulation Line (±) | 5 % | |||
| Output Ripple & Noise Max | 120 mV | |||
| Efficiency | At 100% load / 25°C | 87 % | ||
| Mechanical / Environment | ||||
| Temperature | Operation | 0° to 50°C | ||
| Storage | -20° to 80°C | |||
| Relative humidity | Operation | 10 to 95%, non-condensing | ||
| Dimensions | Form factor | 1U Medical | ||
| Width | 100 mm | |||
| Deep | 220 mm | |||
| Height | 40,5 mm | |||
| Height units | 1 HE | |||
| MTBF (Mean Time Between Failures) | MTBF at 25°C | 100.000 h | ||
| Calculation model | MIL-HDBK- 217 | |||
| Certificates / Conformities | ||||
| CE | Yes | |||
| UL | Yes | |||
| FCC | Class B | Yes | ||
| Additional certificates / compliance approvals | 80 PLUS Gold, CB, EN 55011, EN 55022, EN 61000-3-2, EN 61000-3-3, EN 61000-4-2 (Electrostatic Discharge), EN 61000-4-3, EN 61000-4-4 (EFT), EN 61000-4-5 (2 kV Surge Protection), EN 61000-4-6, EN 61000-4-8, EN 61000-4-11, IEC 60601-1-2:2014, VCCI | |||
| RoHS compliant | Yes | |||